Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know (2024)

Article Navigation

Volume 39 Issue 5 1 September 2004

Article Contents

  • Abstract

  • A Rational Approach to Disinfection and Sterilization

  • Current Issues in Disinfection and Sterilization

  • Conclusion

  • Acknowledgments

  • References

  • < Previous

Journal Article

,

William A. Rutala

Hospital Epidemiology, University of North Carolina Health Care System, and Division of Infectious Diseases, University of North Carolina School of Medicine

,

Chapel Hill

Reprints or correspondence: Dr. William A. Rutala, Div. of Infectious Diseases, 130 Mason Farm Rd., Bioinformatics, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599-7030 (brutala@unch.unc.edu).

Search for other works by this author on:

Oxford Academic

David J. Weber

Hospital Epidemiology, University of North Carolina Health Care System, and Division of Infectious Diseases, University of North Carolina School of Medicine

,

Chapel Hill

Search for other works by this author on:

Oxford Academic

Clinical Infectious Diseases, Volume 39, Issue 5, 1 September 2004, Pages 702–709, https://doi.org/10.1086/423182

Published:

01 September 2004

Article history

Received:

15 March 2004

Accepted:

05 May 2004

Published:

01 September 2004

  • PDF
  • Split View
  • Views
    • Article contents
    • Figures & tables
    • Video
    • Audio
    • Supplementary Data
  • Cite

    Cite

    William A. Rutala, David J. Weber, Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know, Clinical Infectious Diseases, Volume 39, Issue 5, 1 September 2004, Pages 702–709, https://doi.org/10.1086/423182

    Close

Search

Close

Search

Advanced Search

Search Menu

Abstract

All invasive procedures involve contact between a medical device or surgical instrument and a patient's sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of pathogenic microbes that could lead to infection. Failure to properly disinfect or sterilize reusable medical equipment carries a risk associated with breach of the host barriers. The level of disinfection or sterilization is dependent on the intended use of the object: critical items (such as surgical instruments, which contact sterile tissue), semicritical items (such as endoscopes, which contact mucous membranes), and noncritical items (such as stethoscopes, which contact only intact skin) require sterilization, high-level disinfection, and low-level disinfection, respectively. Cleaning must always precede high-level disinfection and sterilization. Users must consider the advantages and disadvantages of specific methods when choosing a disinfection or sterilization process. Adherence to these recommendations should improve disinfection and sterilization practices in health care facilities, thereby reducing infections associated with contaminated patient-care items.

In 1996 in the United States, ∼46,500,000 surgical procedures and an even larger number of invasive medical procedures were performed [1]. For example, ∼5 million gastrointestinal endoscopies are performed per year [1]. Each of these procedures involves contact by a medical device or surgical instrument with a patient's sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of pathogenic microbes, which can lead to infection. For example, failure to properly disinfect or sterilize equipment may lead to person-to-person transmission via contaminated devices (e.g., Mycobacterium tuberculosis—contaminated bronchoscopes).

Achieving disinfection and sterilization through the use of disinfectants and sterilization practices is essential for ensuring that medical and surgical instruments do not transmit infectious pathogens to patients. Because it is not necessary to sterilize all patient-care items, health care policies must identify whether cleaning, disinfection, or sterilization is indicated, primarily on the basis of each item's intended use.

Multiple studies in many countries have documented lack of compliance with established guidelines for disinfection and sterilization [2, 3]. Failure to comply with scientifically based guidelines has led to numerous outbreaks of infection [3–7]. In this article, a pragmatic approach to the judicious selection and proper use of disinfection and sterilization processes is presented that is based on the results of well-designed studies assessing the efficacy (via laboratory investigations) and effectiveness (via clinical studies) of disinfection and sterilization procedures.

A Rational Approach to Disinfection and Sterilization

More than 35 years ago, Spaulding [8] devised a rational approach to disinfection and sterilization of patient-care items or equipment. This classification scheme is so clear and logical that it has been retained, refined, and successfully used by infection-control professionals and others when planning methods for disinfection or sterilization [9–15]. Spaulding believed that the nature of disinfection could be understood more readily if instruments and items for patient care were divided into 3 categories—namely, critical, semicritical, and noncritical—on the basis of the degree of risk of infection involved in the use of the items. This terminology is employed by the Centers for Disease Control and Prevention (CDC) in the documents “Guidelines for Environmental Infection Control in Health-Care Facilities” [16] and “Guideline for Disinfection and Sterilization in Healthcare Facilities” [14].

Critical items. Critical items are those associated with a high risk of infection if the item is contaminated with any microorganism, including bacterial spores. Thus, sterilization of objects that enter sterile tissue or the vascular system is critical, because any microbial contamination could result in disease transmission. This category includes surgical instruments, cardiac and urinary catheters, implants, and ultrasound probes used in sterile body cavities. The items in this category should be purchased as sterile or should be sterilized by steam sterilization, if possible. If the item is heat sensitive, it may be treated with ethylene oxide (ETO) or hydrogen peroxide gas plasma or with liquid chemical sterilants if other methods are unsuitable. Tables 1 and 2 list several germicides that are categorized as chemical sterilants. These include ⩾2.4% glutaraldehyde—based formulations, 1.12% glutaraldehyde with 1.93% phenol/phenate, 7.5% stabilized hydrogen peroxide, 7.35% hydrogen peroxide with 0.23% peracetic acid, ⩾0.2% peracetic acid, and 1.0% hydrogen peroxide with 0.08% peracetic acid. The indicated exposure times are within the range 3–12 h, with the exception of ⩾0.2% peracetic acid (sporicidal time of 12 min at 50°C–56°C) [19]. Use of liquid chemical sterilants is a reliable method of sterilization only if cleaning precedes treatment, which eliminates organic and inorganic material, and if the proper guidelines for concentration, contact time, temperature, and pH are followed. Another limitation to sterilization of devices with liquid chemical sterilants is that the devices cannot be wrapped during processing in the liquid chemical sterilant; thus, maintaining sterility after processing and during storage is impossible. Furthermore, after exposure to the liquid chemical sterilant, devices may require rinsing with water that, in general, is not sterile. Therefore, because of the inherent limitations of the use of liquid chemical sterilants in a nonautomated reprocessor, their use should be restricted to reprocessing critical devices that are heat sensitive and incompatible with other sterilization methods.

Table 1

Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know (3)

Open in new tabDownload slide

Methods for disinfection and sterilization of patient-care items and environmental surfaces.

Table 2

Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know (4)

Open in new tabDownload slide

Summary of advantages and disadvantages of chemical agents used as chemical sterilants or as high-level disinfectants.

Semicritical items. Semicritical items are those that come in contact with mucous membranes or nonintact skin. Respiratory-therapy and anesthesia equipment, some endoscopes, laryngoscope blades, esophageal manometry probes, anorectal manometry catheters, and diaphragm-fitting rings are included in this category. These medical devices should be free of all microorganisms (i.e., mycobacteria, fungi, viruses, and bacteria), although small numbers of bacterial spores may be present. In general, intact mucous membranes, such as those of the lungs or the gastrointestinal tract, are resistant to infection by common bacterial spores but are susceptible to other organisms, such as bacteria, mycobacteria, and viruses. The minimum requirement for semicritical items is high-level disinfection using chemical disinfectants. Glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde (OPA), peracetic acid with hydrogen peroxide, and chlorine have been cleared by the US Food and Drug Administration (FDA) [19] and are dependable high-level disinfectants when guidelines for effective germicidal procedures are followed (tables 1 and 2). The exposure time for most high-level disinfectants varies from 10 to 45 min, at 20°C–25°C. Outbreaks of infection continue to occur when ineffective disinfectants, including iodophor, alcohol, and over-diluted glutaraldehyde [5], are used for so-called high-level disinfection. When a disinfectant is selected for use with certain patient-care items, the chemical compatibility after extended use with the items to be disinfected must also be considered. For example, compatibility testing by Olympus America of 7.5% hydrogen peroxide showed cosmetic and functional changes in the tested endoscopes (Olympus America, personal communication). Similarly, Olympus America does not endorse the use of products containing hydrogen peroxide with peracetic acid, because of cosmetic and functional damage (Olympus America, personal communication).

Semicritical items that will have contact with the mucous membranes of the respiratory or gastrointestinal tract should be rinsed with sterile water, filtered water, or tap water, followed by an alcohol rinse [14, 20, 21]. An alcohol rinse and forced-air drying markedly reduces the likelihood of contamination of the instrument (e.g., endoscopes), most likely by eliminating the wet environment favorable to bacterial growth [21]. After rinsing, items should be dried and then stored in a manner that protects them from damage or contamination. There is no recommendation to use sterile or filtered water, rather than tap water, for rinsing semicritical equipment that will have contact with the mucous membranes of the rectum (e.g., rectal probes or anoscopes) or vagin* (e.g., vagin*l probes) [14].

Noncritical items. Noncritical items are those that come in contact with intact skin but not mucous membranes. Intact skin acts as an effective barrier to most microorganisms; therefore, the sterility of items coming in contact with intact skin is “not critical.” Examples of noncritical items are bedpans, blood-pressure cuffs, crutches, bed rails, linens, bedside tables, patient furniture, and floors. In contrast to critical and some semicritical items, most noncritical reusable items may be decontaminated where they are used and do not need to be transported to a central processing area. There is virtually no documented risk of transmitting infectious agents to patients via noncritical items [22] when they are used as noncritical items and do not contact nonintact skin and/or mucous membranes. However, these items (e.g., bedside tables or bed rails) could potentially contribute to secondary transmission, by contaminating the hands of health care workers or by contact with medical equipment that will subsequently come in contact with patients [23]. Table 1 lists several low-level disinfectants that may be used for noncritical items. The exposure times for these disinfectants are 60 s or longer.

Current Issues in Disinfection and Sterilization

Reprocessing of endoscopes. Physicians use endoscopes to diagnose and treat numerous medical disorders. Although endoscopes are a valuable diagnostic and therapeutic tool in modern medicine and although the incidence of infection associated with their use has been reported to be very low (∼1 in 1.8 million procedures) [24], more health care—associated outbreaks of infection have been linked to contaminated endoscopes than to any other medical device [3–5]. To prevent the spread of health care—associated infection, all heat-sensitive endoscopes (e.g., gastrointestinal endoscopes, bronchoscopes, and nasopharyngoscopes) must be properly cleaned and, at a minimum, subjected to high-level disinfection after each use. High-level disinfection can be expected to destroy all microorganisms, although a few bacterial spores may survive when high numbers of spores are present.

Recommendations for the cleaning and disinfection of endoscopic equipment have been published and should be strictly followed [14, 20]. Unfortunately, audits have shown that personnel do not adhere to guidelines on reprocessing [25–27] and that outbreaks of infection continue to occur [28, 29]. To ensure that the personnel responsible for reprocessing are properly trained, initial and annual competency testing should be required for each individual who is involved in reprocessing endoscopic instruments [14, 20, 21, 30].

In general, endoscope disinfection or sterilization with a liquid chemical sterilant or high-level disinfectant involves the following 5 steps, which should be performed after leak testing: (1) clean: mechanically clean internal and external surfaces, including brushing internal channels and flushing each internal channel with water and an enzymatic cleaner; (2) disinfect: immerse endoscope in high-level disinfectant (or chemical sterilant), perfuse disinfectant (which eliminates air pockets and ensures contact of the germicide with the internal channels) into all accessible channels, such as the suction/biopsy channel and the air/water channel, and expose endoscope for the time recommended for specific products; (3) rinse: rinse the endoscope and all channels with sterile water, filtered water (commonly used with automated endoscope reprocessors), or tap water; (4) dry: rinse the insertion tube and inner channels with alcohol and dry with forced air, after disinfection and before storage; and (5) store: store the endoscope in a way that prevents recontamination and promotes drying (e.g., hung vertically).

Unfortunately, there is poor compliance with the recommendations for reprocessing endoscopes. In addition, in rare instances, the scientific literature and recommendations from professional organizations regarding the use of disinfectants and sterilants may differ from claims on the manufacturer's label. One example is the contact time used to achieve high-level disinfection with 2% glutaraldehyde. On the basis of FDA requirements (the FDA regulates liquid sterilants and high-level disinfectants used on critical and semicritical medical devices), manufacturers test the efficacy of their germicide formulations under worst-case conditions (i.e., minimum recommended concentration of the active ingredient) and in the presence of organic soil (typically, 5% serum). The soil represents the organic loading to which the device is exposed during actual use and that would remain on the device in the absence of cleaning. These stringent test conditions are designed to provide a margin of safety, by assuring that the contact conditions for the germicide provide complete elimination of the test bacteria (e.g., 105–106 cfu M. tuberculosis in organic soil and dried on a scope) if inoculated into the most difficult areas for the disinfectant to penetrate and in the absence of cleaning. However, the scientific data demonstrate that M. tuberculosis levels can be reduced by at least 8 log10 cfu with cleaning (reduction of 4 log10 cfu) followed by chemical disinfection for 20 min at 20°C (reduction of 4–6 log10 cfu) [14, 15, 19, 20, 31]. Because of these data, professional organizations (at least 14 worldwide) that have endorsed an endoscope-reprocessing guideline recommend contact with 2% glutaraldehyde for 20 min (or <20 min outside the United States) at 20°C to achieve high-level disinfection, which differs from the recommendation given on the manufacturer's label [20, 32–34].

It is important to emphasize that the FDA tests do not include cleaning, a critical component of the disinfection process. When cleaning has been included in the test methodology, contact with 2% glutaraldehyde for 20 min has been demonstrated to be effective in eliminating all vegetative bacteria.

Inactivation of Creutzfeldt-Jakob disease (CJD) agent. CJD is a degenerative neurologic disorder in humans, with an incidence in the United States of ∼1 case/million population/year [35]. CJD is thought to be caused by a proteinaceous infectious agent, or prion. CJD is related to other human transmissible spongiform encephalopathies (TSEs), such as kuru (now eradicated), Gertsmann-Straussler-Sheinker syndrome (1 case/40 million population/year), and fatal insomnia syndrome (<1 case/40 million population/year). The agents of CJD and other TSEs exhibit an unusual resistance to conventional chemical and physical decontamination methods. Because the CJD agent is not readily inactivated by conventional disinfection and sterilization procedures and because of the invariably fatal outcome of CJD, the procedures for disinfection and sterilization of the CJD prion have been both conservative and controversial for many years.

The current recommendations consider inactivation data but also use epidemiological studies of prion transmission, infectivity of human tissues, and efficacy of removing proteins by cleaning. On the basis of scientific data, only critical devices (e.g., surgical instruments) and semicritical devices contaminated with high-risk tissue (i.e., brain, spinal cord, or eye tissue) from high-risk patients (e.g., known or suspected infection with CJD or other prion disease) require special prion reprocessing. When high-risk tissues, high-risk patients, and critical or semicritical medical devices are involved, one of the following methods should be used: cleaning of the device and sterilization using a combination of sodium hydroxide and autoclaving [36] (e.g., immerse in 1N NaOH for 1 h, remove and rinse in water, and then transfer to an open pan for autoclaving for 1 h [at 121°C in a gravity displacement sterilizer or at 134°C in a porous or prevacuum sterilizer]); autoclaving for 18 min at 134°C in a prevacuum sterilizer; or autoclaving for 1 h at 132°C in a gravity displacement sterilizer) [14, 37]. The temperature should not exceed 134°C, because the effectiveness of autoclaving may decline as the temperature is increased (e.g., to 136°C or 138°C) [38]. Prion-contaminated medical devices that are impossible or difficult to clean should be discarded. Flash sterilization (i.e., steam sterilization of an unwrapped item for 3 min at 132°C) should not be used for reprocessing. To minimize environmental contamination, noncritical environmental surfaces should be covered with plastic-backed paper; when contaminated with high-risk tissues, the paper should be properly discarded. Noncritical environmental surfaces (e.g., laboratory surfaces) contaminated with high-risk tissues should be cleaned and then spot decontaminated with a 1:10 dilution of hypochlorite solution [37].

Emerging pathogens, antibiotic-resistant bacteria, and bioterrorism agents. Emerging pathogens are of growing concern to the general public and infection-control professionals. Relevant pathogens include Cryptosporidium parvum, Helicobacter pylori, Escherichia coli O157:H7, HIV, hepatitis C virus, rotavirus, multidrug-resistant M. tuberculosis, human papillomavirus, and nontuberculosis mycobacteria (e.g., Mycobacterium chelonae). Similarly, recent publications have highlighted concern about the potential for biological terrorism [39]. The CDC has categorized several agents as “high priority” because they can be easily disseminated or transmitted by person-to-person contact, can cause high mortality, and are likely to cause public panic and social disruption [40]. These agents include Bacillus anthracis (anthrax), Yersinia pestis (plague), variola major (smallpox), Francisella tularensis (tularemia), filoviruses (Ebola and Marburg [hemorrhagic fever]), and arenaviruses (Lassa [Lassa fever] and Junin [Argentine hemorrhagic fever]) and related viruses [40].

With rare exceptions (e.g., human papillomavirus), the susceptibility of each of these pathogens to chemical disinfectants or sterilants has been studied, and all of these pathogens (or surrogate microbes, such as feline calicivirus for Norwalk virus, vaccinia for variola [41], and Bacillus atrophaeus [formerly Bacillus subtilis] for B. anthracis) have been found to be susceptible to currently available chemical disinfectants or sterilants [42]. Standard sterilization and disinfection procedures for patient-care equipment (as recommended in this article) are adequate for sterilization or disinfection of instruments or devices contaminated with blood or other body fluids from persons infected with bloodborne pathogens, emerging pathogens, or bioterrorism agents, with the exception of prions (see previous section). No changes in procedures for cleaning, disinfecting, or sterilizing need to be made [14, 15]. In addition, there are no data to show that antibiotic-resistant bacteria (e.g., methicillin-resistant Staphylococcus aureus, vancomycin-resistant Enterococcus faecium, and multidrug-resistant M. tuberculosis) are less sensitive to liquid chemical germicides than are antibiotic-sensitive bacteria at currently used germicide contact conditions and concentrations [15, 43, 44].

Advances in disinfection and sterilization methods. In the past several years, new methods of disinfection and sterilization have been introduced in health care settings. OPA is a chemical sterilant that received FDA clearance in October 1999. It contains 0.55% 1,2-benzenedicarboxaldehyde. In vitro studies have demonstrated excellent microbicidal activity [14, 15]. For example, Gregory et al. [45] demonstrated that OPA has shown superior mycobactericidal activity (reduction of 5 log10 in 5 min), when compared with glutaraldehyde. The advantages, disadvantages, and characteristics of OPA are listed in table 2 [15].

The FDA recently cleared a liquid high-level disinfectant (superoxidized water) that contains 650–675 ppm free chlorine and a new sterilization system using ozone. Because there are limited data in the scientific literature for assessing the antimicrobial activity or material compatibility of these processes, they have not yet been integrated into clinical practice in the United States [14].

Several methods are used to sterilize patient-care items in health care, including steam sterilization, ETO, hydrogen peroxide gas plasma, and a peracetic acid—immersion system. The advantages and disadvantages of these systems are listed in table 3 [14].

Table 3

Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know (5)

Open in new tabDownload slide

Summary of advantages and disadvantages of commonly used sterilization technologies.

New sterilization technology based on plasma was patented in 1987 and has been marketed in the United States since 1993. Gas plasmas have been referred to as the fourth state of matter (i.e., liquid, solid, gas, and gas plasma). Gas plasmas are generated in an enclosed chamber in a deep vacuum, by using radio frequency or microwave energy to excite the gas molecules and produce charged particles, many of which are in the form of free radicals. This process has the ability to inactivate a broad spectrum of microorganisms, including resistant bacterial spores. Studies have been conducted against vegetative bacteria (including mycobacteria), yeasts, fungi, viruses, and bacterial spores [14]. The effectiveness of all sterilization processes can be altered by lumen length, lumen diameter, inorganic salts, and organic materials [14].

Conclusion

When properly used, disinfection and sterilization can ensure the safe use of invasive and noninvasive medical devices. The method of disinfection and sterilization depends on the intended use of the medical device: critical items (those that contact sterile tissue) must be sterilized prior to use; semicritical items (those that contact mucous membranes or nonintact skin) must undergo high-level disinfection; and noncritical items (those that contact intact skin) should undergo low-level disinfection. Cleaning should always precede high-level disinfection and sterilization. Current disinfection and sterilization guidelines must be strictly followed.

Acknowledgments

Conflict of interest. W.A.R.: Honoraria from Advanced Sterilization Products; consultant for Advanced Sterilization Products, Clorox, 3M, and Metrex; and research funding from Clorox and Reckitt-Benckiser. D.J.B.: Honoraria from Clorox.

References

1

Centers for Disease Control and Prevention (CDC)

,

Ambulatory and inpatient procedures in the United States, 1996

,

1998

Atlanta

CDC

(pg.

1

-

39

)

Google Scholar

OpenURL Placeholder Text

2

McCarthy

GM

,

Koval

JJ

,

John

MA

,

MacDonald

JK

.

Infection control practices across Canada: do dentists follow the recommendations?

,

J Can Dent Assoc

,

1999

,vol.

65

(pg.

506

-

11

)

Google Scholar

OpenURL Placeholder Text

3

Spach

DH

,

Silverstein

FE

,

Stamm

WE

.

Transmission of infection by gastrointestinal endoscopy and bronchoscopy

,

Ann Intern Med

,

1993

,vol.

118

(pg.

117

-

28

)

4

Weber

DJ

,

Rutala

WA

.

Lessons from outbreaks associated with bronchoscopy

,

Infect Control Hosp Epidemiol

,

2001

,vol.

22

(pg.

403

-

8

)

5

Weber

DJ

,

Rutala

WA

,

DiMarino

AJ

Jr

.

DiMarino

AJ

Jr

,

Benjamin

SB

.

The prevention of infection following gastrointestinal endoscopy: the importance of prophylaxis and reprocessing

,

Gastrointestinal diseases: an endoscopic approach

,

2002

Thorofare, NJ

Slack

(pg.

87

-

106

)

Google Scholar

OpenURL Placeholder Text

6

Meyers

H

,

Brown-Elliott

BA

,

Moore

D

,et al.

An outbreak of Mycobacterium chelonae infection following liposuction

,

Clin Infect Dis

,

2002

,vol.

34

(pg.

1500

-

7

)

7

Lowry

PW

,

Jarvis

WR

,

Oberle

AD

,et al.

Mycobacterium chelonae causing otitis media in an ear-nose-and-throat practice

,

N Engl J Med

,

1988

,vol.

319

(pg.

978

-

82

)

8

Spaulding

EH

.

Lawrence

C

,

Block

SS

.

Chemical disinfection of medical and surgical materials

,

Disinfection, sterilization, and preservation

,

1968

Philadelphia

Lea & Febiger

(pg.

517

-

31

)

9

Favero

MS

,

Bond

WW

.

Block

SS

.

Chemical disinfection of medical and surgical materials

,

Disinfection, sterilization, and preservation

,

2001

Philadelphia

Lippincott, Williams & Wilkins

(pg.

881

-

917

)

Google Scholar

OpenURL Placeholder Text

10

Simmons

BP

.

CDC guidelines for the prevention and control of nosocomial infections: guideline for hospital environmental control

,

Am J Infect Control

,

1983

,vol.

11

(pg.

97

-

120

)

11

Garner

JS

,

Favero

MS

.

CDC guideline for handwashing and hospital environmental control, 1985

,

Infect Control

,

1986

,vol.

7

(pg.

231

-

43

)

12

Rutala

WA

.

APIC guideline for selection and use of disinfectants

,

Am J Infect Control

,

1990

,vol.

18

(pg.

99

-

117

)

13

Rutala

WA

.

APIC guideline for selection and use of disinfectants. Association for Professionals in Infection Control and Epidemiology Guidelines Committee of 1994, 1995, and 1996

,

Am J Infect Control

,

1996

,vol.

24

(pg.

313

-

42

)

14

Rutala

WA

,

Weber

DJ

.

Guideline for disinfection and sterilization in healthcare facilities: recommendations of the CDC. Healthcare Infection Control Practices Advisory Committee

,

MMWR Morb Mortal Wkly Rep (in press)

15

Rutala

WA

,

Weber

DJ

.

Mayhall

CG

.

Selection and use of disinfectants in healthcare

,

Hospital epidemiology and infection control

,

2004

Philadelphia

Lippincott, Williams & Wilkins

(pg.

1473

-

522

)

Google Scholar

OpenURL Placeholder Text

16

Sehulster

L

,

Chinn

RYW

.

Guidelines for environmental infection control in health-care facilities. Healthcare Infection Control Practices Advisory Committee

,

MMWR Recomm Rep

,

2003

,vol.

52

(pg.

1

-

44

)

Google Scholar

OpenURL Placeholder Text

17

Kohn

WG

,

Collins

AS

,

Cleveland

JL

,

Harte

JA

,

Eklund

KJ

,

Malvitz

DM

.

Guidelines for infection control in dental health-care settings—2003

,

MMWR Recomm Rep

,

2003

,vol.

52

(pg.

1

-

67

)

Google Scholar

OpenURL Placeholder Text

18

Rutala

WA

,

Weber

DJ

.

Disinfection of endoscopes: review of new chemical sterilants used for high-level disinfection

,

Infect Control Hosp Epidemiol

,

1999

,vol.

20

(pg.

69

-

76

)

19

Food and Drug Administration (FDA)

,

FDA-cleared sterilants and high level disinfectants with general claims for processing reusable medical and dental devices, November

,

2003

Available at: http://www.fda.gov/cdrh/ode/germlab.html. Accessed 8 August 2004

Google Scholar

OpenURL Placeholder Text

20

Nelson

DB

,

Jarvis

WR

,

Rutala

WA

,et al.

Multi-society guideline for reprocessing flexible gastrointestinal endoscopes

,

Infect Control Hosp Epidemiol

,

2003

,vol.

24

(pg.

532

-

7

)

21

Gerding

DN

,

Peterson

LR

,

Vennes

JA

.

Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying

,

Gastroenterology

,

1982

,vol.

83

(pg.

613

-

8

)

Google Scholar

OpenURL Placeholder Text

22

Weber

DJ

,

Rutala

WA

.

Wenzel

RP

.

Environmental issues and nosocomial infections

,

Prevention and control of nosocomial infections

,

1997

Baltimore

Williams & Wilkins

(pg.

491

-

514

)

Google Scholar

OpenURL Placeholder Text

23

Weber

DJ

, .

Role of environmental contamination in the transmission of vancomycin-resistant enterococci

,

Infect Control Hosp Epidemiol

,

1997

,vol.

18

(pg.

306

-

9

)

24

Schembre

DB

.

Infectious complications associated with gastrointestinal endoscopy

,

Gastrointest Endosc Clin N Am

,

2000

,vol.

10

(pg.

215

-

32

)

Google Scholar

OpenURL Placeholder Text

25

Jackson

FW

,

Ball

MD

.

Correction of deficiencies in flexible fiberoptic sigmoidoscope cleaning and disinfection technique in family practice and internal medicine offices

,

Arch Fam Med

,

1997

,vol.

6

(pg.

578

-

82

)

26

Orsi

GB

,

Filocamo

A

,

Di Stefano

L

,

Tittobello

A

.

Italian national survey of digestive endoscopy disinfection procedures

,

Endoscopy

,

1997

,vol.

7

quiz

(pg.

739

-

40

)

Google Scholar

OpenURL Placeholder Text

27

Honeybourne

D

,

Neumann

CS

.

An audit of bronchoscopy practice in the United Kingdom: a survey of adherence to national guidelines

,

Thorax

,

1997

,vol.

52

(pg.

709

-

13

)

28

Srinivasan

A

,

Wolfenden

LL

,

Song

X

,et al.

An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes

,

N Engl J Med

,

2003

,vol.

348

(pg.

221

-

7

)

29

Cetse

JC

,

Vanhems

P

.

Outbreak of infection associated with bronchoscopes

,

N Engl J Med

,

2003

,vol.

348

(pg.

2039

-

40

)

30

Food and Drug Administration (FDA), Centers for Disease Control and Prevention (CDC)

,

FDA and CDC public health advisory: infections from endoscopes inadequately reprocessed by an automated endoscope reprocessing system

,

1999

Rockville, MD

FDA

Google Scholar

OpenURL Placeholder Text

31

Rutala

WA

,

Weber

DJ

.

FDA labeling requirements for disinfection of endoscopes: a counterpoint

,

Infect Control Hosp Epidemiol

,

1995

,vol.

16

(pg.

231

-

5

)

32

Kruse

A

,

Rey

JF

.

Guidelines on cleaning and disinfection in GI endoscopy: update 1999. The European Society of Gastrointestinal Endoscopy

,

Endoscopy

,

2000

,vol.

32

(pg.

77

-

80

)

Google Scholar

OpenURL Placeholder Text

33

British Society of Gastroenterology

Cleaning and disinfection of equipment for gastrointestinal endoscopy: report of a working party of the British Society of Gastroenterology Endoscope Committee

,

Gut

,

1998

,vol.

42

(pg.

585

-

93

)

34

British Thoracic Society

British Thoracic Society guidelines on diagnostic flexible bronchoscopy

,

Thorax

,

2001

,vol.

56

(pg.

1

-

21

)

35

Centers for Disease Control Prevention

Surveillance for Creutzfeldt-Jakob disease—United States

,

MMWR Morb Mortal Wkly Rep

,

1996

,vol.

45

(pg.

665

-

8

)

OpenURL Placeholder Text

36

World Health Organization (WHO)

,

WHO infection control guidelines for transmissible spongiform encephalopathies

37

Rutala

WA

,

Weber

DJ

.

Creutzfeldt-Jakob disease: recommendations for disinfection and sterilization

,

Clin Infect Dis

,

2001

,vol.

32

(pg.

1348

-

56

)

38

Taylor

DM

.

Inactivation of prions by physical and chemical means

,

J Hosp Infect

,

1999

,vol.

43

Suppl

(pg.

69

-

76

)

39

Henderson

DA

.

The looming threat of bioterrorism

,

Science

,

1999

,vol.

283

(pg.

1279

-

82

)

40

Centers for Disease Control Prevention

Biological and chemical terrorism: strategic plan for preparedness and response

,

MMWR Recomm Rep

,

2000

,vol.

49

(pg.

1

-

14

)

OpenURL Placeholder Text

41

Klein

M

,

DeForest

A

.

The inactivation of viruses by germicides

,

Chem Specialists Manuf Assoc Proc

,

1963

,vol.

49

(pg.

116

-

8

)

Google Scholar

OpenURL Placeholder Text

42

Rutala

WA

,

Weber

DJ

.

Infection control: the role of disinfection and sterilization

,

J Hosp Infect

,

1999

,vol.

43

Suppl

(pg.

43

-

55

)

43

Rutala

WA

,

Stiegel

MM

,

Sarubbi

FA

,

Weber

DJ

.

Susceptibility of antibiotic-susceptible and antibiotic-resistant hospital bacteria to disinfectants

,

Infect Control Hosp Epidemiol

,

1997

,vol.

18

(pg.

417

-

21

)

44

Weber

DJ

,

Rutala

WA

.

Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance?

,

Infect Control Hosp Epidemiol (in press)

45

Gregory

AW

,

Schaalje

GB

,

Smart

JD

,

Robison

RA

.

The mycobactericidal efficacy of ortho-phthalaldehyde and the comparative resistances of Mycobacterium bovis, Mycobacterium terrae, and Mycobacterium chelonae

,

Infect Control Hosp Epidemiol

,

1999

,vol.

20

(pg.

324

-

30

)

46

Rutala

WA

,

Weber

DJ

.

Clinical effectiveness of low-temperature sterilization technologies

,

Infect Control Hosp Epidemiol

,

1998

,vol.

19

(pg.

798

-

804

)

© 2004 by the Infectious Diseases Society of America

Topic:

  • disinfection
  • endoscopes
  • patient care
  • surgical procedures, operative
  • infections
  • mucous membrane
  • medical devices
  • microorganisms

Issue Section:

Invited Article

Download all slides

Comments

0 Comments

Comments (0)

Submit a comment

You have entered an invalid code

Thank you for submitting a comment on this article. Your comment will be reviewed and published at the journal's discretion. Please check for further notifications by email.

Advertisem*nt

Citations

Views

131,653

Altmetric

More metrics information

Metrics

Total Views 131,653

122,674 Pageviews

8,979 PDF Downloads

Since 12/1/2016

Month: Total Views:
December 2016 7
January 2017 58
February 2017 203
March 2017 185
April 2017 78
May 2017 102
June 2017 114
July 2017 115
August 2017 145
September 2017 135
October 2017 111
November 2017 161
December 2017 1,116
January 2018 1,319
February 2018 1,712
March 2018 2,339
April 2018 2,904
May 2018 2,004
June 2018 1,525
July 2018 1,353
August 2018 2,096
September 2018 1,912
October 2018 1,770
November 2018 1,843
December 2018 1,956
January 2019 1,787
February 2019 2,124
March 2019 2,297
April 2019 2,130
May 2019 2,761
June 2019 2,253
July 2019 2,213
August 2019 1,972
September 2019 2,572
October 2019 3,066
November 2019 2,472
December 2019 1,710
January 2020 1,854
February 2020 2,191
March 2020 2,792
April 2020 3,609
May 2020 1,623
June 2020 2,212
July 2020 2,178
August 2020 1,713
September 2020 2,091
October 2020 2,544
November 2020 2,403
December 2020 2,116
January 2021 1,782
February 2021 2,385
March 2021 3,472
April 2021 2,797
May 2021 1,922
June 2021 1,426
July 2021 1,368
August 2021 1,404
September 2021 1,591
October 2021 1,819
November 2021 1,883
December 2021 1,359
January 2022 1,300
February 2022 1,385
March 2022 1,583
April 2022 1,276
May 2022 1,317
June 2022 1,088
July 2022 1,605
August 2022 858
September 2022 1,088
October 2022 1,253
November 2022 1,066
December 2022 957
January 2023 997
February 2023 1,108
March 2023 1,211
April 2023 1,013
May 2023 882
June 2023 956
July 2023 824
August 2023 781
September 2023 856
October 2023 874
November 2023 734
December 2023 612
January 2024 694
February 2024 876
March 2024 1,305

Citations

Powered by Dimensions

191 Web of Science

Altmetrics

×

Email alerts

Article activity alert

Advance article alerts

New issue alert

Receive exclusive offers and updates from Oxford Academic

More on this topic

Rapid Environmental Contamination With Candida auris and Multidrug-Resistant Bacterial Pathogens Near Colonized Patients

Effectiveness of Ultraviolet-C Disinfection on Hospital-Onset Gram-Negative Rod Bloodstream Infection: A Nationwide Stepped-Wedge Time-Series Analysis

Ultraviolet-C Light Evaluation as Adjunct Disinfection to Remove Multidrug-Resistant Organisms

Control of Multidrug-Resistant Pseudomonas aeruginosa in Allogeneic Hematopoietic Stem Cell Transplant Recipients by a Novel Bundle Including Remodeling of Sanitary and Water Supply Systems

Citing articles via

Google Scholar

  • Latest

  • Most Read

  • Most Cited

Revolutionizing Daptomycin Dosing: A Single 7-11 Hour Sample for Pragmatic Application
True DisCoVeRy of COVID-19 Disease Burden Versus Speculated Antiviral Cardiovascular Risk Requires a Control Group
Cardiac Adverse Events and Remdesivir in Hospitalized Patients with Coronavirus Disease 2019 (COVID-19): A Post Hoc Safety Analysis of the Randomized DisCoVeRy Trial
A Phase 2 Clinical Trial to Evaluate the Safety, Reactogenicity, and Immunogenicity of Different Prime-Boost Vaccination Schedules of 2013 and 2017 A(H7N9) Inactivated Influenza Virus Vaccines Administered with and without AS03 Adjuvant in Healthy US Adults
Epidemiology and Treatment Outcomes of Tuberculosis with Chronic Hepatitis B Infection—California, 2016–2020

More from Oxford Academic

Clinical Medicine

Infectious Diseases

Medicine and Health

Books

Journals

Advertisem*nt

Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know (2024)
Top Articles
Latest Posts
Article information

Author: Rev. Leonie Wyman

Last Updated:

Views: 5974

Rating: 4.9 / 5 (59 voted)

Reviews: 82% of readers found this page helpful

Author information

Name: Rev. Leonie Wyman

Birthday: 1993-07-01

Address: Suite 763 6272 Lang Bypass, New Xochitlport, VT 72704-3308

Phone: +22014484519944

Job: Banking Officer

Hobby: Sailing, Gaming, Basketball, Calligraphy, Mycology, Astronomy, Juggling

Introduction: My name is Rev. Leonie Wyman, I am a colorful, tasty, splendid, fair, witty, gorgeous, splendid person who loves writing and wants to share my knowledge and understanding with you.